  • People with medical implants (such as cardiac implants pacemakers, aneurysm clips, surgical clips or prostheses) should check with us before entering the NMR room.
  •  
  • All magnetic objects should be kept outside the NMR magnet room.
  •  
  • Keep electronics, credit cards and magnetic storage media outside of the NMR magnet room.
  •  
  • In case a metallic object strikes the magnet, get NMR Facility staff immediately. Do NOT attempt to pull the object off yourself.
  •  
  • In the event of a magnet quench (sudden and rapid boil-off of the magnets cryogens (helium and nitrogen) that can be detected by visible (and/or audible) emission of cryogenic gas from the magnet), the magnet cryogens that vaporizes displace air and can result in asphyxiation. If a quench occurs, immediately evacuate the NMR room!
 